[121714085] (b)(4). Investigation results: a stone cone retrieval coil was returned for analysis. A visual evaluation found a part of the coil was detached and a burn mark at the break. The device opens and closes. Additionally, the handle was found broken into two pieces. The reported event that the device could not close could not be confirmed; however there is a burn mark at the break. The stone cone directions for use warns: do not directly fire upon the device with a laser. Therefore, the assigned complaint investigation conclusion code for this event is user / use error. Confirmed through complaint investigation that there was an act or omission of an act that resulted in a different medical device response than intended by the manufacturer or expected by the user. A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.
Patient Sequence No: 1, Text Type: N, H10


[121714086] It was reported to boston scientific corporation that a stone cone was used in a rigid ureteroscopy procedure performed on (b)(6) 2018. According to the complainant, during the procedure, the stone cone failed to straighten after it coiled. The procedure was completed with a different device. There were no patient complications reported as a result of this event. The patients' condition at the conclusion of the procedure was reported to be stable. This has been deemed an mdr reportable event based on investigation findings coil/coating peeled/shared/frayed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-02608
MDR Report Key7906508
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-09-25
Date of Report2018-09-14
Date of Event2018-06-21
Date Mfgr Received2018-09-14
Device Manufacturer Date2018-01-24
Date Added to Maude2018-09-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTONE CONE NITINOL UROLOGICAL RETRIEVAL COIL
Generic NameDISLODGER, STONE, FLEXIBLE
Product CodeFGO
Date Received2018-09-25
Returned To Mfg2018-07-24
Model NumberM0063903100
Catalog Number390-310
Lot Number0004123922
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - MARLBOROUGH
Manufacturer Address100 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.