MAUDE MDR 7909689

MDR report key
7909689
Report number
1319808-2018-00025
Event key
0
Event type
3
Date of event
2018-08-29
Date received
2018-09-26
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
MR. JAMES A STEVENS
Address
100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US
Phone
585-585-5854
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1VITROS CHEMISTRY PRODUCTS CALIBRATOR KIT 2IN VITRO DIAGNOSTICSORTHO-CLINICAL DIAGNOSTICSJIX16626590277Y Y

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12018-09-260

Event Narratives#

N

Patient 1

THE INVESTIGATION DETERMINED THAT HIGHER THAN EXPECTED VITROS K+ AND NA+ RESULTS WERE OBTAINED FROM A NON-VITROS BIORAD LEVEL 1 QUALITY CONTROL FLUID PROCESSED USING VITROS K+ AND NA+ SLIDES IN COMBINATION WITH A VITROS 5600 INTEGRATED SYSTEM. THE EVENT WAS ISOLATED TO THE CALIBRATION EVENTS PERFORMED ON 29 AND 30 AUGUST 2018 USING THE SAME CALIBRATOR KIT LOT 0277. THE CALIBRATIONS WERE SUBOPTIMAL WHEN COMPARED TO THE TYPICAL CALIBRATION RESPONSES AND PARAMETERS. THE MOST LIKELY CAUSE OF THE SUBOPTIMAL CALIBRATIONS WAS USER ERROR. THE CUSTOMER SUSPECTED THE CALIBRATOR FLUIDS IN USE WERE NOT PROPERLY RECONSTITUTED DUE TO USER ERROR, ALTHOUGH THIS COULD NOT BE DEFINITIVELY CONFIRMED. ACCEPTABLE VITROS NA+ AND K+ PERFORMANCE WAS OBSERVED AFTER A RECALIBRATION EVENT WAS PERFORMED USING PROPERLY HANDLED CALIBRATOR FLUIDS.

D

Patient 1

THE CUSTOMER OBTAINED HIGHER THAN EXPECTED VITROS K+ AND NA+ RESULTS FROM A NON-VITROS BIORAD LEVEL 1 QUALITY CONTROL FLUID PROCESSED USING VITROS K+ AND NA+ SLIDES IN COMBINATION WITH A VITROS 5600 INTEGRATED SYSTEM. BIORAD LEVEL 1, LOT 47960 NA+ RESULTS OF 143.0, 142.7, AND 158.1 MMOL/L VS. THE EXPECTED RESULT OF 115.0 MMOL/L. BIORAD LEVEL 1, LOT 47960 K+ RESULT OF 3.70 MMOL/L VS. THE EXPECTED RESULT OF 2.50 MMOL/L. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE UNEXPECTED VITROS NA+ AND K+ RESULTS WERE GENERATED FROM NON-PATIENT FLUIDS, HOWEVER THE INVESTIGATION CANNOT CONCLUDE THAT PATIENT SAMPLE RESULTS WERE NOT AFFECTED AND WOULD NOT BE AFFECTED IF THE EVENT WERE TO RECUR UNDETECTED. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THE EVENT. THIS REPORT IS NUMBER ONE OF TWO MDR?S FOR THIS EVENT. TWO 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS TWO DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. COMPLAINT NUMBERS (B)(4).