MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-09-26 for VITEK? 2 GRAM NEGATIVE IDENTIFICATION TEST KIT 21341 manufactured by Biomerieux Inc..
[122309750]
An industry customer from the united states notified biom? Rieux of a misidentification result for the moraxella osloensis atcc 19976 strain when testing with the vitek? 2 gn id card (ref 21341). The gn card identified the sample as acinetobacter lowffii on initial testing. As there is no patient associated with this quality control strain, there is no adverse event related to any patient's state of health. An internal biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1950204-2018-00393 |
MDR Report Key | 7912363 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2018-09-26 |
Date of Report | 2018-12-10 |
Date Mfgr Received | 2018-11-13 |
Device Manufacturer Date | 2018-07-30 |
Date Added to Maude | 2018-09-26 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. CANDACE MARTIN |
Manufacturer Street | 595 ANGLUM ROAD |
Manufacturer City | HAZELWOOD MO 63042 |
Manufacturer Country | US |
Manufacturer Postal | 63042 |
Manufacturer G1 | BIOMERIEUX INC. |
Manufacturer Street | 595 ANGLUM ROAD |
Manufacturer City | SAINT LOUIS MO 63042 |
Manufacturer Country | US |
Manufacturer Postal Code | 63042 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | VITEK? 2 GRAM NEGATIVE IDENTIFICATION TEST KIT |
Generic Name | VITEK? 2 GN ID TEST KIT |
Product Code | LQM |
Date Received | 2018-09-26 |
Catalog Number | 21341 |
Lot Number | 2410611203 |
Device Expiration Date | 2019-07-30 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BIOMERIEUX INC. |
Manufacturer Address | 595 ANGLUM ROAD SAINT LOUIS MO 63042 US 63042 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-09-26 |