HEMOCHRON SIGNATURE ELITE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-09-26 for HEMOCHRON SIGNATURE ELITE manufactured by Accriva Diagnostics.

Event Text Entries

[122163884] This mdr references accriva diagnostics' complaint (b)(4) for the hemochron signature elite instrument, serial (b)(4). Complaint (b)(4) captures the child case for the lot number of the jact+ cuvette used during the procedure. Accriva diagnostics has requested all data required for form 3500a. This report serves as the initial and final mdr for this case and effectively closes the complaint.
Patient Sequence No: 1, Text Type: N, H10


[122163885] A healthcare professional reported inconsistent act results with a hemochron signature elite and act+ whole blood microcoagulation system during an ablation procedure for an abnormal heart rhythm. The target act for this procedure was 350 seconds. The third act reading was 258 seconds, which was a lower than expected result. The same blood sample was re-tested on a different elite instrument. The act reading was 370 seconds, which was within the expected range and was used for patient management. The procedure was completed successfully and no adverse events were reported. Both electronic and liquid quality controls passed on the suspect instrument prior to the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2250033-2018-00022
MDR Report Key7912705
Report SourceHEALTH PROFESSIONAL
Date Received2018-09-26
Date of Report2018-08-27
Date of Event2018-08-13
Date Mfgr Received2018-08-27
Device Manufacturer Date2016-05-01
Date Added to Maude2018-09-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJON MCDERMED
Manufacturer Street6260 SEQUENCE DRIVE
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8582632490
Manufacturer G1ACCRIVA DIAGNOSTICS
Manufacturer Street6260 SEQUENCE DRIVE
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal Code92121
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHEMOCHRON SIGNATURE ELITE
Generic NameSYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES
Product CodeJPA
Date Received2018-09-26
Returned To Mfg2018-09-07
Model NumberELITE
Catalog NumberELITE
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerACCRIVA DIAGNOSTICS
Manufacturer Address6260 SEQUENCE DRIVE SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.