CLINICAL CHEMISTRY CARBON DIOXIDE 03L80-31

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-09-27 for CLINICAL CHEMISTRY CARBON DIOXIDE 03L80-31 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[123533587] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. An evaluation is in-process.
Patient Sequence No: 1, Text Type: N, H10


[123533588] The customer observed falsely elevated co2 results while using the clinical chemistry co2 assay. The customer provided the following data for 4 patients (normal range is 23-31 meq/l): sid (b)(6): initial 32 meq/l, retested using new reagent 27 meq/l. Sid (b)(6): initial 35 meq/l, retested using new reagent 29 meq/l. Sid (b)(6): initial 32 meq/l, retested using new reagent 27 meq/l. Sid (b)(6): initial 31 meq/l, retested using new reagent 27 meq/l. No impact to patient management was reported. No additional patient information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2018-01916
MDR Report Key7914125
Report SourceHEALTH PROFESSIONAL
Date Received2018-09-27
Date of Report2019-01-25
Date of Event2018-09-12
Date Mfgr Received2019-01-03
Device Manufacturer Date2018-03-28
Date Added to Maude2018-09-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINICAL CHEMISTRY CARBON DIOXIDE
Generic NameCARBON DIOXIDE
Product CodeKHS
Date Received2018-09-27
Catalog Number03L80-31
Lot Number52577UQ02
Device Expiration Date2019-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-27

© 2025 FDA.report
This site is not affiliated with or endorsed by the FDA.