GRIP-LOK SMALL HYDROCOLLOID SECUREMENT 3200S

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-09-27 for GRIP-LOK SMALL HYDROCOLLOID SECUREMENT 3200S manufactured by Tidi Products, Llc.

Event Text Entries

[122008241] Evaluation codes: since the product was not returned for evaluation, the reported issue could not be confirmed. This report is being submitted solely on the reported issue. If the device does not adhere as expected, catheter could become dislodged or risk of infection if there is not a good seal. This report is being submitted as a result of a remediation capa to reassess product risks. No further actions will be taken. All complaints are trended and reviewed by management on a monthly basis. As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[122008242] Customer reported the device is not holding the catheter on a child as long as it should. Customer believes this may be causing lateral tension and possible occlusion of the lines. It is unknown when the issue was discovered and no patient incident or injury was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2182318-2018-00013
MDR Report Key7914735
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-09-27
Date of Report2018-09-24
Date of Event2017-06-06
Date Mfgr Received2018-09-24
Device Manufacturer Date2015-02-17
Date Added to Maude2018-09-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJACKIE KRIEGLSTEIN
Manufacturer Street570 ENTERPRISE DRIVE
Manufacturer CityNEENAH WI 54956
Manufacturer CountryUS
Manufacturer Postal54956
Manufacturer Phone9207514300
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGRIP-LOK SMALL HYDROCOLLOID SECUREMENT
Generic NameGRIP-LOK 3200S
Product CodeKMK
Date Received2018-09-27
Model Number3200S
Catalog Number3200S
Lot Number16004807
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTIDI PRODUCTS, LLC
Manufacturer Address570 ENTERPRISE DRIVE NEENAH WI 54956 US 54956


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.