ABL80 FLEX ANALYZER 393-839

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2018-09-27 for ABL80 FLEX ANALYZER 393-839 manufactured by Sendx Medical, Inc..

Event Text Entries

[121914791] A patient analysis was performed on the subject analyzer, abl80 flex serial number (b)(4), and based on the reported hct/thb result the patient received a blood transfusion. The user was suspicious of the subject analyzer's result and therefore performed another measurement on a comparison abl80 flex that was on-site. Measured results: subject analyzer: thb = 7. 1g/dl hct = 22%. Comparison analyzer: thb = 13. 7 g/dl hct = 42%. Based on a comparison of the results between the subject and comparison analyzers, the user concluded that the subject analyzer results were incorrect and the blood transfusion was unnecessary. A service technician evaluated the analyzer and found that the roller wheel assembly was not working properly. Further investigation revealed the roller wheel assembly housing was cracked around a set screw that is used to secure the roller wheel assembly to the roller motor shaft. The cracked housing resulted in roller wheel assembly slippage affecting sample aspiration and flushing. The issue was resolved following replacement of the roller wheel assembly.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027541-2018-00002
MDR Report Key7914865
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2018-09-27
Date of Report2018-09-26
Date of Event2018-09-10
Date Mfgr Received2018-09-10
Device Manufacturer Date2006-10-31
Date Added to Maude2018-09-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ANGIE HSU
Manufacturer Street1945 PALOMAR WAY
Manufacturer CityCARLSBAD CA 92011
Manufacturer CountryUS
Manufacturer Postal92011
Manufacturer Phone7608033481
Manufacturer G1SENDX MEDICAL, INC.
Manufacturer Street1945 PALOMAR WAY
Manufacturer CityCARLSBAD CA 92011
Manufacturer CountryUS
Manufacturer Postal Code92011
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameABL80 FLEX ANALYZER
Generic NameELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH
Product CodeJPI
Date Received2018-09-27
Model Number393-839
Catalog Number393-839
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSENDX MEDICAL, INC.
Manufacturer Address1945 PALMOAR OAKS WAY CARLSBAD CA 92011 US 92011


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-09-27