CHROMID? CARBA SMART 414685

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-09-28 for CHROMID? CARBA SMART 414685 manufactured by Biomerieux S.a..

Event Text Entries

[124861446] A customer in (b)(6) notified biomerieux of no growth and delayed results associated with chromid? Carba smart (reference (b)(4)). The customer reported observing no growth on chromid for escherichia coli on the oxa 48 side and small growth on the carba side of the plate. The antibiogram of the strain was resistant and the customer sent it to a reference laboratory. At the reference laboratory e. Coli was confirmed with the presence of a carbapenamse oxa 48 and the presence of an extended-spectrum beta-lactamases (esbl). Patient treatment was not impacted by this discrepancy. Initial treatment of the patient occurred in (b)(6) with bactrim (r), which did not relieve the patient's complaints for one (1) week. The patient sought further treatment in (b)(6). The patient's condition required nitrofurantoin according to the antibiogram, which he received in (b)(6), but hospitalization was not required. The following results were reported to the physician: "sent to the reference lab due to suspected cpe. " after one (1) week the confirmation by the reference lab was reported to the doctor: "cpe" (carbapenem-producing enterobacteriaceae), carbapenemase type oxa-48. Results were delayed due to the discrepant results by two (2) to three (3) days. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00189
MDR Report Key7918708
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-09-28
Date of Report2018-12-17
Date Mfgr Received2018-11-19
Device Manufacturer Date2018-06-18
Date Added to Maude2018-09-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. TIFFANY HALL
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer G1BIOMERIEUX S.A.
Manufacturer Street5 RUE DES AQUEDUCS 692290 CRAPONNE
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID? CARBA SMART
Generic NameCHROMID? CARBA SMART
Product CodeJSO
Date Received2018-09-28
Catalog Number414685
Lot Number1006583300
Device Expiration Date2018-09-02
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX S.A.
Manufacturer Address5 RUE DES AQUEDUCS CRAPONNE, 69290 FR 69290


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-28

© 2026 FDA.report