MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-10-05 for TRIVEX SYSTEM CONTROL UNIT WITH BUILT IN PUMP 7210387 manufactured by Lemaitre Vascular, Inc..
[123413068]
We have received the device for evaluation. The blades rotated properly when the run button was engaged. The blades rotated intermittently and hesitantly when the window lock button was engaged. Upon disassembly of the housing, we noticed water in the motor and pc board. The probable root cause of the defect is a damage to the motor and the pc board due to water exposure. We currently have a corrective and preventive action (capa) open to address this issue. The device was not used in the patient. The issue was detected during pre-use check.
Patient Sequence No: 1, Text Type: N, H10
[123413069]
During pre-use check, the blades of the handpiece rotated intermittently.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1220948-2018-00075 |
MDR Report Key | 7939095 |
Date Received | 2018-10-05 |
Date of Report | 2018-10-05 |
Date of Event | 2018-09-06 |
Date Mfgr Received | 2018-09-07 |
Date Added to Maude | 2018-10-05 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. PRAGYA THIKEY |
Manufacturer Street | 63 SECOND AVE |
Manufacturer City | BURLINGTON MA 01803 |
Manufacturer Country | US |
Manufacturer Postal | 01803 |
Manufacturer Phone | 7812212266 |
Manufacturer G1 | LEMAITRE VASCULAR, INC. |
Manufacturer Street | 63 SECOND AVE |
Manufacturer City | BURLINGTON MA 01803 |
Manufacturer Country | US |
Manufacturer Postal Code | 01803 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | TRIVEX SYSTEM CONTROL UNIT WITH BUILT IN PUMP |
Generic Name | HANDPIECE |
Product Code | DWQ |
Date Received | 2018-10-05 |
Returned To Mfg | 2018-09-14 |
Catalog Number | 7210387 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | LEMAITRE VASCULAR, INC. |
Manufacturer Address | 63 SECOND AVE BURLINGTON MA 02148 US 02148 |