RUTNER SUPRAPUBIC BALLOON CATHETER SET G15261 081912

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,o report with the FDA on 2018-10-08 for RUTNER SUPRAPUBIC BALLOON CATHETER SET G15261 081912 manufactured by Cook Inc.

Event Text Entries

[123203886] (b)(6). (b)(4). This report includes information known at this time. A follow up report will be submitted should additional relevant information become available.
Patient Sequence No: 1, Text Type: N, H10


[123203887] It was reported the rutner suprapubic balloon catheter was placed for percutaneous cystostomy, but the urine drainage was not confirmed next day. The customer found the catheter easily came off and the balloon was deflated. To find out what was the cause, he put the balloon into the water and attempted to inflate and confirmed the balloon would not inflate. He replaced it with another balloon. The patient did not experience any adverse effects due to this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2018-03026
MDR Report Key7942447
Report SourceFOREIGN,HEALTH PROFESSIONAL,O
Date Received2018-10-08
Date of Report2018-11-06
Date of Event2018-09-25
Date Mfgr Received2018-10-31
Device Manufacturer Date2018-01-16
Date Added to Maude2018-10-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. LARRY POOL
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRUTNER SUPRAPUBIC BALLOON CATHETER SET
Generic NameKOB CATHETER, SUPRAPUBIC (AND ACCESSORIES)
Product CodeKOB
Date Received2018-10-08
Returned To Mfg2018-10-08
Model NumberG15261
Catalog Number081912
Lot NumberNS8513221
Device Expiration Date2019-07-16
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-08

© 2026 FDA.report