MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-10-09 for VITEK? 2 GP (GRAM-POSITIVE) ID TEST KIT 21342 manufactured by Biomerieux, Inc..
[123948453]
A customer in (b)(6) reported a misidentification of staphylococcus aureus as staphylococcus pseudintermedius in association with the vitek? 2 gp (gram-positive) id test kit (lot 2420686403). The customer reported that the gp id card identified s. Pseudintermedius and both vitek ms and api? Identified s. Aureus. The test was repeated on two vitek 2 instruments and the results were s. Pseudintermedius. Both instruments were recently serviced and passed qc. The customer stated the patient result was not affected and there was a 24 hour delay due to repeat testing. There is no indication or report from the hospital or treating physician to biom? Rieux that the discrepant result led to any adverse event related to the patient's state of health. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1950204-2018-00417 |
MDR Report Key | 7949735 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2018-10-09 |
Date of Report | 2018-12-17 |
Date Mfgr Received | 2018-11-20 |
Device Manufacturer Date | 2018-04-13 |
Date Added to Maude | 2018-10-09 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. DEBRA BROYLES |
Manufacturer Street | 595 ANGLUM ROAD |
Manufacturer City | HAZELWOOD MO 63042 |
Manufacturer Country | US |
Manufacturer Postal | 63042 |
Manufacturer G1 | BIOMERIEUX, INC. |
Manufacturer Street | 595 ANGLUM ROAD |
Manufacturer City | HAZELWOOD MO 63042 |
Manufacturer Country | US |
Manufacturer Postal Code | 63042 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | VITEK? 2 GP (GRAM-POSITIVE) ID TEST KIT |
Generic Name | VITEK? 2 GP (GRAM-POSITIVE) ID TEST KIT |
Product Code | LQL |
Date Received | 2018-10-09 |
Catalog Number | 21342 |
Lot Number | 2420686403 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BIOMERIEUX, INC. |
Manufacturer Address | 595 ANGLUM ROAD HAZELWOOD MO 63042 US 63042 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-10-09 |