VITEK? 2 GP (GRAM-POSITIVE) ID TEST KIT 21342

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-10-09 for VITEK? 2 GP (GRAM-POSITIVE) ID TEST KIT 21342 manufactured by Biomerieux, Inc..

Event Text Entries

[123948453] A customer in (b)(6) reported a misidentification of staphylococcus aureus as staphylococcus pseudintermedius in association with the vitek? 2 gp (gram-positive) id test kit (lot 2420686403). The customer reported that the gp id card identified s. Pseudintermedius and both vitek ms and api? Identified s. Aureus. The test was repeated on two vitek 2 instruments and the results were s. Pseudintermedius. Both instruments were recently serviced and passed qc. The customer stated the patient result was not affected and there was a 24 hour delay due to repeat testing. There is no indication or report from the hospital or treating physician to biom? Rieux that the discrepant result led to any adverse event related to the patient's state of health. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1950204-2018-00417
MDR Report Key7949735
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-10-09
Date of Report2018-12-17
Date Mfgr Received2018-11-20
Device Manufacturer Date2018-04-13
Date Added to Maude2018-10-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DEBRA BROYLES
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOMERIEUX, INC.
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal Code63042
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITEK? 2 GP (GRAM-POSITIVE) ID TEST KIT
Generic NameVITEK? 2 GP (GRAM-POSITIVE) ID TEST KIT
Product CodeLQL
Date Received2018-10-09
Catalog Number21342
Lot Number2420686403
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, INC.
Manufacturer Address595 ANGLUM ROAD HAZELWOOD MO 63042 US 63042


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.