SHOE DH OFFLOAD MD 10341

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-10-09 for SHOE DH OFFLOAD MD 10341 manufactured by Ossur Americas.

Event Text Entries

[123195722] According to the patient, the velcro hook broke and became undone, which caused a fall and an injury to the patient's ankle. The patient mentioned that he did seek medical attention in form of x-ray. The product was not returned, and it is difficult to pinpoint the failure based on the customer description as there are two hook and loop straps on the product. Therefore, there are a variety of factors that must be considered regarding why the hook didn't adhere properly to the loop. These factors may include, shoe size too small, loop worn out due to repeated use, user error during donning. This is a newly identified failure; therefore, no action is required at this moment. Attempts have been made to contact the customer for additional information regarding what, if any, injury occurred. No reply from customer. Customer service followed up to get patient status on many occasions with no response from patient. Dates follow up was performed: (b)(6) 2018, (b)(6) 2018 - (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[123195723] Patient claimed he fell, when the velcro failure on the dh offloading post-op shoe and had to seek medical attention in form of an xray for his ankle.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2085446-2018-00003
MDR Report Key7949768
Date Received2018-10-09
Date of Report2018-07-17
Date Mfgr Received2018-07-17
Date Added to Maude2018-10-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag0
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS CRYSTAL SWAFFORD
Manufacturer StreetNO. 11, TSU-CHIANG 1ST ROAD,
Manufacturer CityNANTOU CITY NANTOU, TAIWAN 540
Manufacturer CountryTW
Manufacturer Postal540
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Sequence Number: 0

Brand NameSHOE DH OFFLOAD MD
Generic NameSHOE DH OFFLOAD MD
Product CodeIPG
Date Received2018-10-09
Model Number10341
Catalog Number10341
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerOSSUR AMERICAS
Manufacturer Address27051 TOWNE CENTRE DRIVE FOOTHILL RANCH CA 92610 US 92610

Device Sequence Number: 1

Brand NameSHOE DH OFFLOAD MD
Generic NameSHOE DH OFFLOAD MD
Product CodeIPG
Date Received2018-10-09
Model Number10341
Catalog Number10341
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOSSUR AMERICAS
Manufacturer Address27051 TOWNE CENTRE DRIVE FOOTHILL RANCH CA 92610 US 92610


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2018-10-09