MALIBU/SOVEREIGN AZL33216-GB

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2018-10-10 for MALIBU/SOVEREIGN AZL33216-GB manufactured by Arjo Hospital Equipment Ab.

Event Text Entries

[123690536] (b)(4). The collected information is under further review. Additional information will be provided within the next report.
Patient Sequence No: 1, Text Type: N, H10


[123690537] It was reported that nurse received an electric shock from the lift arm, when trying to move it manually. No patient was involved. There is no indication of any injury or medical treatment. The device was taken out of use and inspected by arjo representative. The fault could have not been replicated. The bath was tested for an electrical safety on all metal parts and it passed all tests. It was suspected that this was a static electric shock.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007420694-2018-00196
MDR Report Key7951287
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2018-10-10
Date of Report2018-10-31
Date of Event2018-09-21
Date Facility Aware2018-09-24
Report Date2018-10-31
Date Reported to FDA2018-10-31
Date Reported to Mfgr2018-10-31
Date Mfgr Received2018-09-24
Device Manufacturer Date2009-10-26
Date Added to Maude2018-10-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KINGA STOLINSKA
Manufacturer StreetKS. WAWRZYNIAKA 2
Manufacturer CityKOMORNIKI, 62-052
Manufacturer CountryPL
Manufacturer Postal62-052
Manufacturer G1ARJO HOSPITAL EQUIPMENT AB
Manufacturer StreetVERKSTADSVAGEN 5
Manufacturer CityESLOV, 24121
Manufacturer CountrySW
Manufacturer Postal Code24121
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMALIBU/SOVEREIGN
Generic NameBATH, HYDRO-MASSAGE
Product CodeILJ
Date Received2018-10-10
Model NumberAZL33216-GB
Device AvailabilityY
Device Age9 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerARJO HOSPITAL EQUIPMENT AB
Manufacturer AddressVERKSTADSVAGEN 5 ESLOV, 24121 SW 24121


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-10-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.