ARCHITECT TOTAL PSA 07K70-30

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2018-10-10 for ARCHITECT TOTAL PSA 07K70-30 manufactured by Abbott Ireland.

Event Text Entries

[125464902] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. This report is being filed on an international product, list number 7k70 that has a similar product distributed in the us, list number 6c06. An evaluation is in process.
Patient Sequence No: 1, Text Type: N, H10


[125464903] The customer observed falsely depressed results while using architect total psa (prostate specific antigen) reagents. The following data was provided for the same patient. Sid (b)(6) initial 0. 83 ng/ml, repeat 0. 80, 0. 79, 0. 74 ng/ml, repeat in another laboratory 0. 71, 0. 61, 0. 83, 0. 88 ng/ml. Sid (b)(6) initial 1. 19 ng/ml, not repeated. Sid (b)(6) initial 2. 02 ng/ml, not repeated. Sid (b)(6) initial 1. 16 ng/ml, not repeated. Sid (b)(6) initial 3. 38 ng/ml, not repeated. Sid (b)(6) initial 1. 09 ng/ml, not repeated. Sid (b)(6) initial 4. 57 ng/ml, not repeated. Sid (b)(6) initial 7. 23 ng/ml, not repeated. Sid (b)(6) initial 0. 74 ng/ml, not repeated. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008344661-2018-00085
MDR Report Key7954341
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2018-10-10
Date of Report2018-10-10
Date of Event2018-09-06
Date Mfgr Received2018-09-20
Date Added to Maude2018-10-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT IRELAND
Manufacturer StreetDIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK
Manufacturer CitySLIGO NA
Manufacturer CountryEI
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT TOTAL PSA
Generic NameTOTAL PSA
Product CodeLTJ
Date Received2018-10-10
Catalog Number07K70-30
Lot Number85185FN00
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerABBOTT IRELAND
Manufacturer AddressDIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK SLIGO NA EI NA


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-10