[125936939] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[125936940] The customer complained of a discrepant elecsys anti-ccp immunoassay result for 1 patient sample on cobas 8000 e 602 module serial number (b)(4). The patient had gone to the rheumatologist for pain. The initial anti-ccp result was 213. 3 u/l on the e 602 module. The physician questioned the result and the sample was sent to another laboratory. The anti-ccp result was negative with the pharmacia immunocap method. The anti-ccp result was 0 u/l with the euro diagnostica elisa method. There was no allegation of an adverse event. The rheumatologist stated that patients' anti-ccp results are often positive with roche tests and negative with other methods.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-03560
MDR Report Key7955597
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-10-11
Date of Report2018-11-05
Date of Event2018-09-19
Date Mfgr Received2018-09-24
Date Added to Maude2018-10-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS ANTI-CCP IMMUNOASSAY
Generic NameANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP)
Product CodeNHX
Date Received2018-10-11
Model NumberNA
Catalog Number05031656190
Lot Number34774000
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-11