CAPITAL TEMP PUMP PROFESSIONAL TP700

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-10-12 for CAPITAL TEMP PUMP PROFESSIONAL TP700 manufactured by Stryker Medical-kalamazoo.

Event Text Entries

[123677217] The following event was reported via user facility medwatch (b)(4): it was reported that a patient has 2 one centimeter burn wounds presumed to be from the heating pad. One wound is healed and the second is being treated by the wound team.
Patient Sequence No: 1, Text Type: D, B5


[134207112] Attempts were made to contact the user facility to gather further information regarding this alleged issue, however, the customer did not respond to these attempts. A review of device functionality and a conversation with a stryker clinical nurse consultant identified that for a burn to occur, the product would need to be greater than 44 degrees celsius. Without a failure of the backup and primary thermostats, the highest temperature the t-pump can reach is 42 degrees celsius. As such, this issue could not have occurred without a failure of both the primary and backup limit thermostats. This could not be confirmed. Device not returned.
Patient Sequence No: 1, Text Type: N, H10


[134207113] The following event was reported via user facility medwatch (b)(4): it was reported that a patient has 2 one centimeter burn wounds presumed to be from the heating pad. One wound is healed and the second is being treated by the wound team.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001831750-2018-01092
MDR Report Key7962627
Date Received2018-10-12
Date of Report2019-01-10
Date of Event2018-08-01
Date Mfgr Received2018-09-17
Date Added to Maude2018-10-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag0
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BRIAN THOMPSON
Manufacturer Street3800 EAST CENTRE AVENUE
Manufacturer CityPORTAGE MI 49002
Manufacturer CountryUS
Manufacturer Postal49002
Manufacturer Phone2693292100
Manufacturer G1STRYKER MEDICAL-KALAMAZOO
Manufacturer Street3800 EAST CENTRE AVENUE
Manufacturer CityPORTAGE MI 49002
Manufacturer CountryUS
Manufacturer Postal Code49002
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCAPITAL TEMP PUMP PROFESSIONAL
Generic NamePACK, HOT OR COLD, WATER CIRCULATING
Product CodeILO
Date Received2018-10-12
Catalog NumberTP700
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER MEDICAL-KALAMAZOO
Manufacturer Address3800 EAST CENTRE AVENUE PORTAGE MI 49002 US 49002


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.