NIM? EMG ELECTRODE 8227410

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-10-15 for NIM? EMG ELECTRODE 8227410 manufactured by Medtronic Xomed Inc..

Event Text Entries

[123745628] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[123745749] A health care professional (hcp) reported they were having issues with their electrodes. When going into their electrode check, the status would stay green. Surgeon was no longer able to get a response on the system, it would just constantly beep as if it was constantly stimulating. They turned the channel off and the beeping went away, removed the electrodes and turned the channel back on, electrodes off was displayed, indicating that the interface box was working seeing the electrodes just fine. This confirmed that the nim was isolated with no other systems or chords close to it. The hcp rebooted the system and disconnected and reconnected the interface box. The leads were checked on the patient that they were still in place. The electrodes were unplugged and replugged into the interface box without resolution. The hcp decided to continue without the system. The procedure was in less than a hour. There was no patient impact.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1045254-2018-00513
MDR Report Key7966254
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-10-15
Date of Report2019-02-08
Date of Event2018-09-18
Date Mfgr Received2019-01-28
Date Added to Maude2018-10-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactURIZA SHUMS
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328405
Manufacturer G1MEDTRONIC XOMED INC.
Manufacturer Street6743 SOUTHPOINT DR N
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNIM? EMG ELECTRODE
Generic NameELECTRODE, NEEDLE
Product CodeGXZ
Date Received2018-10-15
Model Number8227410
Catalog Number8227410
Lot NumberNI
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED INC.
Manufacturer Address6743 SOUTHPOINT DR N JACKSONVILLE FL 32216 US 32216


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-15

© 2026 FDA.report