ROD INTRODUCTION PLIERS FOR SIDE-OPENING IMPLANTS 388.50

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-10-15 for ROD INTRODUCTION PLIERS FOR SIDE-OPENING IMPLANTS 388.50 manufactured by Wrights Lane Synthes Usa Products Llc.

Event Text Entries

[123806306] There was no known reported patient involvement associated with the complained event. Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received. A review of the device history records has been requested. The investigation could not be completed; no conclusion could be drawn, as no product was received. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[123806307] It was reported that on an unknown date, a rod introduction plier for side-opening implant tube was rotated and the sleeve (holds the collar implant) that slides up and down the "tube" got jammed on the tube during a routine inspection of consignment instrument sets. The set screw that keeps the sleeve in an open track on the tube was too short and does not go into the track far enough. There was no patient and procedure involvement. This report is for one (1) rod introduction pliers for side-opening implants. This is report 1 of 1 for complaint (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2939274-2018-54387
MDR Report Key7967044
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-10-15
Date of Report2018-10-01
Date Mfgr Received2018-11-28
Device Manufacturer Date1999-01-28
Date Added to Maude2018-10-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MICHAEL COTE
Manufacturer Street1302 WRIGHT LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone6107195000
Manufacturer G1MONUMENT
Manufacturer Street1101 SYNTHES AVENUE
Manufacturer CityMONUMENT CO 80132
Manufacturer CountryUS
Manufacturer Postal Code80132
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameROD INTRODUCTION PLIERS FOR SIDE-OPENING IMPLANTS
Generic NamePLIERS, SURGICAL
Product CodeHTC
Date Received2018-10-15
Returned To Mfg2018-10-23
Model Number388.50
Catalog Number388.50
Lot Number3914478
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerWRIGHTS LANE SYNTHES USA PRODUCTS LLC
Manufacturer Address1302 WRIGHTS LANE EAST WEST CHESTER PA 19380 US 19380


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-15