HUDSON STYLET,SOFT TIP,6 FR 5-15120

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-10-16 for HUDSON STYLET,SOFT TIP,6 FR 5-15120 manufactured by Teleflex Medical.

Event Text Entries

[126118051] Qn# (b)(4). The device involved has not been received by the manufacturer for evaluation at the time of this report. A device history record investigation did not show issues related to this complaint. A record assessment (fmea) was conducted and no update is required. Customer complaint cannot be confirmed due the product sample is not available to perform a proper investigation and determine the root cause. Root cause is unknown. Corrective actions cannot be established. If the device sample becomes available at a later date, this report will be updated accordingly.
Patient Sequence No: 1, Text Type: N, H10


[126118052] Customer complaint alleges "during the intubation of the child, the rigid guide was used with a portex endotracheal tube (manufactured by smith medical) d 2. 5 ref 100/105/025. A part of the guide remained in the tube during removal. (cont. ) the patient was extubated and the detached part of the guide was removed from the young patient. " it was reported there was no consequence for the child. Patient condition reported as fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003898360-2018-00820
MDR Report Key7970816
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-10-16
Date of Report2018-10-01
Date of Event2018-09-29
Date Mfgr Received2018-11-19
Device Manufacturer Date2018-03-08
Date Added to Maude2018-10-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKATHARINE TARPLEY
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334854
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetRANCHO EL DESCANSO
Manufacturer CityTECATE 21478
Manufacturer CountryMX
Manufacturer Postal Code21478
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUDSON STYLET,SOFT TIP,6 FR
Generic NameSTYLET, TRACHEAL TUBE
Product CodeBSR
Date Received2018-10-16
Returned To Mfg2018-10-31
Catalog Number5-15120
Lot Number73C1800199
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-16

© 2026 FDA.report