COC THC OPI AMP MAMP PCP BZO BAR MTD MDM DOA-21107

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-10-17 for COC THC OPI AMP MAMP PCP BZO BAR MTD MDM DOA-21107 manufactured by Alere San Diego, Inc..

Event Text Entries

[124523980] Investigation pending.
Patient Sequence No: 1, Text Type: N, H10


[124523981] The customer reported receiving positive amp, coc, and thc results on two donors using two multi-drug 10 panel cups. The donors denied having drugs in their system. Testing was performed on an alternate lot of multi-drug 10 panel cups and produced negative results. The positive results occurred on 2 devices out of 4 devices used. No laboratory confirmation testing was performed and no additional information available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2018-00138
MDR Report Key7974963
Report SourceHEALTH PROFESSIONAL
Date Received2018-10-17
Date of Report2018-10-17
Date Mfgr Received2018-12-20
Date Added to Maude2018-10-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOC THC OPI AMP MAMP PCP BZO BAR MTD MDM
Generic NameENZYME IMMUNOASSAY, COCAINE AND COCAINE METABOLITES
Product CodeDIO
Date Received2018-10-17
Returned To Mfg2018-11-28
Model NumberDOA-21107
Lot NumberDOA7100507
Device Expiration Date2019-10-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-17

© 2026 FDA.report