DEVICE, PERCUTANEOUS, BIOPSY VPB11

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-10-18 for DEVICE, PERCUTANEOUS, BIOPSY VPB11 manufactured by Medtronic Mexico S. De R.l De C.v..

Event Text Entries

[124528180] Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation. Therefore, we are unable to determine the definitive cause of the reported event. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[124528181] It was reported that bone marrow biopsy was being performed on a patient with hard bone. Intra-op, plastic hub of the biopsy device came off. When the biopsy device was seated to take the specimen, it stuck in the patient. When trying to remove the device, the plastic top of the device began to spin around the biopsy shaft and the plastic top came off; leaving the shaft stuck in the patient. Finally, the shaft was removed with the help of vice grips and the procedure was completed. No fragment of the broken product remained inside the patient. No patient complications were reported due to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612164-2018-02815
MDR Report Key7977964
Date Received2018-10-18
Date of Report2018-10-18
Date of Event2018-09-25
Date Mfgr Received2018-09-25
Date Added to Maude2018-10-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACIE ZIEMBA
Manufacturer Street1800 PYRAMID PLACE
Manufacturer CityMEMPHIS TN 38132
Manufacturer CountryUS
Manufacturer Postal38132
Manufacturer Phone9013963133
Manufacturer G1MEDTRONIC MEXICO S. DE R.L DE C.V.
Manufacturer StreetAVENIDA PASEO CUCAPAH # 10510 PARQUE INDUSTRIAL EI LAGO
Manufacturer CityTIJUANA,BC
Manufacturer CountryMX
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameDEVICE, PERCUTANEOUS, BIOPSY
Generic NameDEVICE, PERCUTANEOUS, BIOPSY
Product CodeMJG
Date Received2018-10-18
Model NumberNA
Catalog NumberVPB11
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC MEXICO S. DE R.L DE C.V.
Manufacturer AddressAVENIDA PASEO CUCAPAH # 10510 PARQUE INDUSTRIAL EI LAGO TIJUANA,BC MX


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-18

© 2026 FDA.report