BLUE COMFORT BOUFFANT 24 IN 3274

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-10-18 for BLUE COMFORT BOUFFANT 24 IN 3274 manufactured by Hubei Safeway Protective Products.

Event Text Entries

[124310973] Based on the supplier? S investigation, the device history record review did not indicate any exception that could lead to the reported incident. There was no change in the material composition. One bouffant was returned for visual inspection, there was no abnormality found. Considering the complaint defect category of irritation, the supplier reviewed the raw material and component inspection record and their existing material used for this code 3274, complies to the specification and no exception was detected that could result in this reported issue. The elastic material used for the cap passed bio-compatibility tests required by the regulatory agency for the intended use. Based on the investigation, the root cause could not be determined. The complaint information was provided to the relevant sectors for their awareness. There is no additional action taken at this time, but the supplier will continue to monitor the trend of this type of incident.
Patient Sequence No: 1, Text Type: N, H10


[124310974] The end user-clinician experienced rash from the elastic on bouffant cap on forehead.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1423537-2018-00243
MDR Report Key7981447
Date Received2018-10-18
Date of Report2018-10-18
Date of Event2018-09-12
Date Facility Aware2018-09-21
Report Date2018-09-14
Date Reported to Mfgr2018-09-14
Date Mfgr Received2018-09-21
Date Added to Maude2018-10-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPATRICIA TUCKER
Manufacturer Street3651 BIRCHWOOD DR.
Manufacturer CityWAUKEGAN IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer Phone8478874151
Manufacturer G1HUBEI SAFEWAY PROTECTIVE PRODUCTS
Manufacturer StreetNO. 118, PENCHANG AVENUE
Manufacturer CityXIANTAO HUBEI 433000
Manufacturer CountryCH
Manufacturer Postal Code433000
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBLUE COMFORT BOUFFANT 24 IN
Generic NameCAP, SURGICAL
Product CodeFYF
Date Received2018-10-18
Returned To Mfg2018-10-04
Model Number3274
Catalog Number3274
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHUBEI SAFEWAY PROTECTIVE PRODUCTS
Manufacturer AddressNO. 118, PENCHANG AVENUE XIANTAO HUBEI 433000 CH 433000


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-10-18