ARCHITECT TOTAL PSA 07K70-27

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2018-10-19 for ARCHITECT TOTAL PSA 07K70-27 manufactured by Abbott Ireland.

Event Text Entries

[126107398] Evaluation of the customer issue included a complaint text review, a search for similar complaints, device history review, labeling review, and field data review. Returns were not available. No adverse trend was identified for the customer issue. Device history review did not identify any issues that may have caused the customer issue. Labeling was reviewed and found to be adequate. Abbott link world wide data was used to evaluate historical product field performance. Patient data was analyzed and compared to an established control limit. The evaluation indicated that the patient median result for the complaint lot was comparable to other lots in the field between (b)(6) 2018 to (b)(6) 2018. No unusual reagent lot performance was identified. Based on all available information and abbott diagnostics evaluation, no systemic issue was identified and no product deficiency was found. This report is being filed on an international product, list number (b)(4) that has a similar product distributed in the us, list number (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[126107399] The customer observed falsely elevated results with poor dilution linearity while using architect total psa (prostate specific antigen) reagents. The following data was provided for the same patient. Initial (undiluted) greater than 100 ng/ml, diluted (1:10) greater than 100 ng/ml, diluted (1:20) 26. 8 ng/ml, diluted (1:100) 214. 24 ng/ml. Per the product package insert, specimens with values exceeding 100 ng/ml are flagged with code greater than 100. 000 and may be diluted. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008344661-2018-00089
MDR Report Key7986516
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2018-10-19
Date of Report2018-10-19
Date Mfgr Received2018-09-27
Device Manufacturer Date2018-02-14
Date Added to Maude2018-10-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT IRELAND
Manufacturer StreetDIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK
Manufacturer CitySLIGO NA
Manufacturer CountryEI
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT TOTAL PSA
Generic NameTOTAL PSA
Product CodeMTF
Date Received2018-10-19
Catalog Number07K70-27
Lot Number84190FN00
Device Expiration Date2019-01-09
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT IRELAND
Manufacturer AddressDIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK SLIGO NA EI NA


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.