ROUND FILTERS W/INDICATOR US751

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-10-22 for ROUND FILTERS W/INDICATOR US751 manufactured by Sps Medical Supply Corp.

Event Text Entries

[126228939] Manufacturing site evaluation: investigation indicates the root cause of the reported issue is: chemical indicator ink has a viscosity which is too high for manufacturing equipment. High viscosity ink becomes stuck to applicators, resulting in splatter throughout the process. The current manufacturing process lacks specifications for ink viscosity. Immediate corrections have been put in place at the manufacturing site including manufacture. Increase in quality control sampling of filters to no less then every 30 minutes of manufacture along with enhanced training and visibility of this issue to operators trained in filter manufacture. Investigation of the makeup of the splatter is the same ink as the chemical indicator ink, due to the visible confirmation of ink color transition post-sterilization, and is not a contaminant nor a deficiency which affects the functionality of the product. This issue has been elevated to a capa to eliminate the issue.
Patient Sequence No: 1, Text Type: N, H10


[126228940] It was reported by the healthcare professional to the company sales representative "the staff notice visible dark spots on filter aside from where the indicator dot is placed on the filter. " there was a 1 hour delay in surgery. No patient injury. No mention if additional intervention was needed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2916714-2018-00025
MDR Report Key7991660
Report SourceHEALTH PROFESSIONAL
Date Received2018-10-22
Date of Report2018-10-22
Date of Event2018-09-26
Date Facility Aware2018-09-26
Date Mfgr Received2018-09-26
Date Added to Maude2018-10-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515988
Manufacturer G1SPS MEDICAL SUPPLY CORP
Manufacturer Street6789 W HENRIETTA ROAD
Manufacturer CityRUSH NY 14543
Manufacturer CountryUS
Manufacturer Postal Code14543
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameROUND FILTERS W/INDICATOR
Generic NameSTERILE TECHNOLOGY
Product CodeKCT
Date Received2018-10-22
Model NumberUS751
Catalog NumberUS751
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSPS MEDICAL SUPPLY CORP
Manufacturer Address6789 W HENRIETTA ROAD RUSH NY 14543 US 14543


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-22

© 2024 FDA.report