[124963283] Information was received that during the use of a smiths medical breathing circuit, water droplets stayed in the connector for the hme filter, within the circuit ahead of the connector, and in the gas sampling line. No patient injury resulted.
Patient Sequence No: 1, Text Type: D, B5


[133986092] One photo of the device was received. Upon visual inspection of the photo, water in the circuit was confirmed. While no definitive root cause to the reported issue could be determined, this investigation revealed no intrinsic evidence to suggest a root cause related to manufacturing.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3012307300-2018-01911
MDR Report Key7997381
Date Received2018-10-23
Date of Report2018-11-21
Date of Event2018-09-01
Date Mfgr Received2018-10-25
Date Added to Maude2018-10-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone7633833310
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameADULT DISPOSABLE ANESTHESIA BREATHING CIRCUIT
Generic NameCIRCULATOR, BREATHING-CIRCUIT
Product CodeCAG
Date Received2018-10-23
Returned To Mfg2018-09-28
Catalog NumberCA24B0/4P6/000JP
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC.
Manufacturer Address6000 NATHAN LANE NORTH MINNEAPOLIS MN 55442 US 55442


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-23