I-STAT CREA CARTRIDGE 03P84-25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-10-24 for I-STAT CREA CARTRIDGE 03P84-25 manufactured by Abbott Point Of Care.

Event Text Entries

[126244142] Apoc incident # (b)(4). Apoc labeling was evaluated during the investigation as pertaining to the event. The investigation was completed on 10/24/2018. A review of the device history record confirmed the lot passed finished goods release criteria. Retained cartridge testing met the acceptance criteria found in q04. 01. 003 rev. Ac, appendix 1- product complaint level 2 and level 3 investigation procedure. No deficiency has been identified.
Patient Sequence No: 1, Text Type: N, H10


[126244143] On (b)(6) 2018, abbott point of care was contacted by a customer regarding i-stat crea+ cartridges that yielded a suspected discrepant creatinine result on an (b)(6) year old male cancer patient. There was no additional patient information available at the time of this report. Return product is not available for investigation. (b)(6). Test and sample collection times were not provided. There are no injuries associated with this event. At this time there is no reason to suspect a malfunction exists. The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay. Based on the i-stat results it is suspected that the patient was taking hydroxyurea although the customer reported that hydroxyurea was not listed on medication list. The ordering doctor stated that the patient was on hydroxyurea back in (b)(6) 2018 but not now.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2245578-2018-00262
MDR Report Key8001409
Report SourceHEALTH PROFESSIONAL
Date Received2018-10-24
Date of Report2018-10-24
Date of Event2018-09-28
Date Mfgr Received2018-09-28
Device Manufacturer Date2018-07-04
Date Added to Maude2018-10-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLINDA MACZUSZENKO
Manufacturer Street400 COLLEGE ROAD
Manufacturer CityPRINCETON NJ 08540
Manufacturer CountryUS
Manufacturer Postal08540
Manufacturer Phone6136885949
Manufacturer G1ABBOTT POINT OF CARE CANADA LTD.
Manufacturer Street185 CORKSTOWN ROAD
Manufacturer CityOTTAWA, ONTARIO K2H 8V4
Manufacturer CountryCA
Manufacturer Postal CodeK2H 8V4
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameI-STAT CREA CARTRIDGE
Generic NameCREA CARTRIDGE
Product CodeCGL
Date Received2018-10-24
Model NumberNA
Catalog Number03P84-25
Lot NumberA18185
Device Expiration Date2018-12-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT POINT OF CARE
Manufacturer Address400 COLLEGE ROAD PRINCETON NJ 085406607 US 085406607


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-24

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.