[125158767] Instructions for use (ins 181a rev c) and quick start guide (999-329 rev a) both specifically state in step 4 of removal process: take one of the recommended removal instruments and place the tip ends at the adhesive interface (concave channel) between the motion 3d appliance cuspid pad and the tooth surface. Orient the instrument toward the mesial aspect of the cuspid or bicuspid pad in an occlusal/gingival aspect. Gently squeeze, applying increased continuous pressure, without any twisting or pulling until the bond fails. In this instance, the dentist stated that the sides of the pad of the device was used. In addition, examination of the returned product has evidence of tool marks on the product, not on the adhesive interface.
Patient Sequence No: 1, Text Type: N, H10


[125158768] While the technician was removing carriere motion 3d colors class ii 25mm left - silver, upper left 3 canine fractured. Per the dentist, the following steps were taken: removal of excess adhesive with high speed hand piece; having patient bite on cotton roll; squeezing on each side of pad with bracket removing plier. Device was bonded on (b)(6) 2018 and was being debonded due to the patient being finished with the treatment step.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2081322-2018-00001
MDR Report Key8009692
Report SourceHEALTH PROFESSIONAL
Date Received2018-10-26
Date of Report2018-10-08
Date of Event2018-10-02
Date Mfgr Received2018-10-08
Date Added to Maude2018-10-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMARY PEARMAN
Manufacturer Street1822 ASTON AVENUE
Manufacturer CityCARLSBAD CA 92008
Manufacturer CountryUS
Manufacturer Postal92008
Manufacturer G1ORTHO ORGANIZERS, INC
Manufacturer Street1822 ASTON AVENUE
Manufacturer CityCARLSBAD CA 920087306
Manufacturer CountryUS
Manufacturer Postal Code920087306
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARRIERE MOTION 3D COLORS CLASS II 25MM LEFT - SILVER
Generic NameDISTALIZER
Product CodeEJF
Date Received2018-10-26
Returned To Mfg2018-10-18
Model Number424-925LN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO ORGANIZERS, INC
Manufacturer Address1822 ASTON AVENUE CARLSBAD CA 920087306 US 920087306


Patients

Patient NumberTreatmentOutcomeDate
101. Deathisabilit 2018-10-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.