AXLE INTERSPINOUS FUSION SYSTEM X060-1520 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-10-26 for AXLE INTERSPINOUS FUSION SYSTEM X060-1520 NA manufactured by X-spine Systems, Inc..

Event Text Entries

[125290772] The complainant reported that an axle all in one inserter broke during insertion of an implant. A second inserter was available, and the procedure was completed. There were no known patient complications. Visual assessment of the instrument identified that a subassembly part was sheared off which would prevent the inserter from functioning as intended. The fracture appeared to start on the inside corner of the tip where there is a sharp corner and potential stress riser. A dhr review was performed, and there were no manufacturing anomalies identified. The device met all required specifications prior to being released to distributable inventory.
Patient Sequence No: 1, Text Type: N, H10


[125290773] The complainant reported that an axle all in one inserter broke during insertion of an implant. A second inserter was available, and the procedure was completed. There were no known patient complications.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005031160-2018-00031
MDR Report Key8010522
Date Received2018-10-26
Date of Report2018-10-26
Date Mfgr Received2018-10-05
Device Manufacturer Date2016-04-15
Date Added to Maude2018-10-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CASEY MING
Manufacturer Street664 CRUISER LANE
Manufacturer CityBELGRADE MT 59714
Manufacturer CountryUS
Manufacturer Postal59714
Manufacturer Phone4063880480
Manufacturer G1X-SPINE SYSTEMS, INC.
Manufacturer Street664 CRUISER LANE
Manufacturer CityBELGRADE MT 59714
Manufacturer CountryUS
Manufacturer Postal Code59714
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameAXLE INTERSPINOUS FUSION SYSTEM
Generic NamePLIERS ALL-IN-ONE INSERTER/COMPRESSOR, ASSEMBLY
Product CodeHTC
Date Received2018-10-26
Returned To Mfg2018-10-09
Model NumberX060-1520
Catalog NumberNA
Lot Number584601
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerX-SPINE SYSTEMS, INC.
Manufacturer Address664 CRUISER LANE BELGRADE MT 59714 US 59714


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-10-26

© 2026 FDA.report