VASCULAR PROBES 7081015ES

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-10-29 for VASCULAR PROBES 7081015ES manufactured by Baxter Healthcare - Saint Paul.

Event Text Entries

[125639756] Lot number: this is a summary report for lot numbers sp18c261297227 and sp18d161300710. Three single use devices were received for evaluation. Visual inspection revealed particulate matter on the outside wall of the inner pouch for all three samples. Microscopic inspection revealed that the particulate matter was a fiber. For two units, the size of the particulate matter was found to be within specification for this device (lot# sp18c261297227 and sp18d161300710). The devices were found to be conforming product. For the third sample (lot# sp18d161300710), the size of the particulate matter was found to be out of specification. The reported condition was verified for the third sample. The cause of the condition was not determined. A batch review was conducted for lots sp18c261297227 and sp18d161300710 and there were no deviations found related to this reported condition during the manufacture of these lots. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[125639757] This report summarizes 3 malfunction events. It was reported that particulate matter was found inside the inner pouch of three vascular probes. For all three events, there was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2018-06874
MDR Report Key8014782
Date Received2018-10-29
Date of Report2018-10-29
Date Mfgr Received2018-09-30
Date Added to Maude2018-10-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - SAINT PAUL
Manufacturer Street2575 UNIVERSITY AVE W
Manufacturer CitySAINT PAUL MN 55114
Manufacturer CountryUS
Manufacturer Postal Code55114
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVASCULAR PROBES
Generic NameDILATOR, VESSEL, SURGICAL
Product CodeDWP
Date Received2018-10-29
Model NumberNA
Catalog Number7081015ES
Lot NumberASKU
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - SAINT PAUL
Manufacturer AddressSAINT PAUL MN


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-29

© 2026 FDA.report