SHOCKPULSE SE LITHOTRIPSY PROBE, REUSABLE, 3.40MM SPL-PR340

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2018-10-29 for SHOCKPULSE SE LITHOTRIPSY PROBE, REUSABLE, 3.40MM SPL-PR340 manufactured by Cybersonics Inc..

Event Text Entries

[125350598] The device was not returned to cybersonics for evaluation. The event occurred during reuse of the device. Instructions for use (ifu) of the device recommends 5 reuses. It is not clear how many times the device has been reused. A review of similar reported complaints indicates that this reported phenomenon is likely attributed to excessive force applied on the device, excessive use of the device and/ or excessive bending during the procedure.
Patient Sequence No: 1, Text Type: N, H10


[125350599] The doctor was performing percutaneous nephrolithotomy procedure (pcnl) and during the procedure, probe broke. The doctor removed the probe from the patient and continued the surgery using another probe. No injury or health hazard was reported to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004216443-2018-00004
MDR Report Key8015458
Report SourceDISTRIBUTOR
Date Received2018-10-29
Date of Report2018-10-30
Date of Event2018-08-22
Date Mfgr Received2018-10-16
Device Manufacturer Date2016-11-14
Date Added to Maude2018-10-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMRS. SAMRADNI PATIL
Manufacturer Street5340 FRYLING ROAD SUITE 101
Manufacturer CityERIE PA 165104703
Manufacturer CountryUS
Manufacturer Postal165104703
Manufacturer Phone8149201510
Manufacturer G1CYBERSONICS INC.
Manufacturer Street5340 FRYLING ROAD SUITE 101
Manufacturer CityERIE PA 165104703
Manufacturer CountryUS
Manufacturer Postal Code165104703
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSHOCKPULSE SE LITHOTRIPSY PROBE, REUSABLE, 3.40MM
Generic NameELECTROHYDRAULIC LITHOTRIPTOR.
Product CodeFEO
Date Received2018-10-29
Model NumberSPL-PR340
Lot NumberW1606167
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCYBERSONICS INC.
Manufacturer Address5340 FRYLING ROAD SUITE 101 ERIE PA 165104703 US 165104703


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.