COSEAL 934073

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-10-29 for COSEAL 934073 manufactured by Baxter Healthcare - Hayward.

Event Text Entries

[125603587] The device was received for evaluation in its unopened pouch. The returned device was labeled for single use. Visual inspection revealed white powder scattered inside the pouch of the dsd syringe. A concentrated area of white powder was observed at the transfer port of the dsd syringe. The reported condition was verified. The cause of the condition was not determined. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[125603588] This report summarizes 1 malfunction events. It was reported that white particulate matter was found inside the transfer port of a coseal surgical sealant product. There was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2018-06907
MDR Report Key8015802
Date Received2018-10-29
Date of Report2018-10-29
Date Mfgr Received2018-09-30
Date Added to Maude2018-10-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - HAYWARD
Manufacturer Street2024 W WINTON AVE
Manufacturer CityHAYWARD CA 94545
Manufacturer CountryUS
Manufacturer Postal Code94545
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCOSEAL
Generic NameSEALANT,POLYMERIZING
Product CodeNBE
Date Received2018-10-29
Model NumberNA
Catalog Number934073
Lot NumberHA170722CC
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - HAYWARD
Manufacturer AddressHAYWARD CA


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-29

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.