DRAGONFLY IMAGING CATHETER 100-100-00

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-10-29 for DRAGONFLY IMAGING CATHETER 100-100-00 manufactured by St. Jude Medical Catd.

Event Text Entries

[125430780] After successful antegrade wire crossing in the occluded artery, pre-dilation was performed several times with several balloons. Dissection was apparent in the lad part. A oct catheter was attempted to be used on a distal lesion but was unable to go through. The oct image displayed a dissection. Since blood flow was affected, stenting was performed and the oct catheter was removed. After the stent was placed a second dissection appeared. Another stent was placed for the second dissection. When a post-stent oct assessment was attempted, the oct catheter was unable to pass through the stent so a full stent assessment was not made. The patient was unstable and the procedure was stopped. Hospital policy prohibits the release of patient specific information. (b)(4), pt id (b)(6).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3009600098-2018-00016
MDR Report Key8016571
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-10-29
Date of Report2018-11-14
Date Mfgr Received2018-11-13
Device Manufacturer Date2018-11-13
Date Added to Maude2018-10-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1ST. JUDE MEDICAL
Manufacturer Street4 ROBBINS DRIVE
Manufacturer CityWESTFORD MA 01886
Manufacturer CountryUS
Manufacturer Postal Code01886
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDRAGONFLY IMAGING CATHETER
Generic NameOPTICAL COHERENCE TOMOGRAPHY, INTRAVASCULAR CATHETER
Product CodeORD
Date Received2018-10-29
Model Number100-100-00
Catalog Number100-100-00
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL CATD
Manufacturer Address4 ROBBINS DRIVE WESTFORD MA 01886 US 01886


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2018-10-29

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.