VIDAS? TSH 30400

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-10-29 for VIDAS? TSH 30400 manufactured by Biomerieux Sa.

Event Text Entries

[126520877] A customer in (b)(6) reported a false low result for one healthy patient during a control exam, with two different lots of the vidas? Tsh assay. The customer stated the tsh result from lot 1006127340 was <0. 05. The result was reported to the endocrinologist, and they did not agree with the result, so the patient was sent to another lab for repeat analysis. The repeat result was within normal values. The customer then obtained another sample from the patient and repeated the tsh assay with lot 1006277930, and the result was <0. 05. The customer reported that the patient's t4 and total t4 values were normal. The customer stated that patient treatment was not impacted and there was no delay. There is no indication or report from the hospital or treating physician to biom? Rieux that the discrepant result led to any adverse event related to the patient's state of health. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00226
MDR Report Key8017067
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-10-29
Date of Report2018-12-27
Date Mfgr Received2018-11-30
Device Manufacturer Date2018-02-13
Date Added to Maude2018-10-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMS. DEBRA BROYLES
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280,
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS? TSH
Generic NameVIDAS? TSH
Product CodeKLI
Date Received2018-10-29
Catalog Number30400
Lot Number1006277930
Device Expiration Date2019-01-29
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280, FR


Patients