MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2018-10-29 for CLINICAL CHEMISTRY LACTATE DEHYDROGENASE 02P56-21 manufactured by Abbott Manufacturing Inc.
[126661163]
An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. An evaluation is in process.
Patient Sequence No: 1, Text Type: N, H10
[126661164]
The customer observed imprecise results while using clinical chemistry lactate dehydrogenase reagents. The following data was provided for one patient. Sid (b)(4), specimen 1 initial (drawn in the morning, tested at 5:00 p. M. Using the c16000 analyzer sn c1601627) 388. 6 u/l, repeat 182. 93 u/l, repeat using c8000 analyzer sn (b)(4) 329. 32, 179. 63 u/l specimen 2 initial (drawn in the afternoon, test time not provided using the c16000 analyzer sn (b)(4) 173. 69 multiple replicates for the patient ranged from 159. 57 to 181. 97 u/l, however no draw time or analyzer used were provided. Per the product package insert, the reference range for the assay is 125 to 220 u/l. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1628664-2018-01980 |
MDR Report Key | 8017113 |
Report Source | FOREIGN,HEALTH PROFESSIONAL |
Date Received | 2018-10-29 |
Date of Report | 2018-12-28 |
Date of Event | 2018-10-11 |
Date Mfgr Received | 2018-12-12 |
Date Added to Maude | 2018-10-29 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NOEMI ROMERO-KONDOS, RN BSN |
Manufacturer Street | 100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3 |
Manufacturer City | ABBOTT PARK IL 600643537 |
Manufacturer Country | US |
Manufacturer Postal | 600643537 |
Manufacturer Phone | 224667-512 |
Manufacturer G1 | ABBOTT MANUFACTURING INC |
Manufacturer Street | 1921 HURD DRIVE |
Manufacturer City | IRVING TX 75038 |
Manufacturer Country | US |
Manufacturer Postal Code | 75038 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CLINICAL CHEMISTRY LACTATE DEHYDROGENASE |
Generic Name | LACTATE DEHYDROGENASE |
Product Code | CFJ |
Date Received | 2018-10-29 |
Catalog Number | 02P56-21 |
Lot Number | 99549UN18 |
Operator | HEALTH PROFESSIONAL |
Device Availability | * |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT MANUFACTURING INC |
Manufacturer Address | 1921 HURD DRIVE IRVING TX 75038 US 75038 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-10-29 |