ELECSYS T4 ASSAY 07027885190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-10-30 for ELECSYS T4 ASSAY 07027885190 manufactured by Roche Diagnostics.

Event Text Entries

[125784877] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[125784878] The customer stated that they received erroneous results for one patient sample tested with the elecsys t4 assay on a cobas 8000 e 801 module. The erroneous results were not reported outside of the laboratory. The sample was tested three times on the customer's e 801 analyzer, resulting with t4 values of 24. 4 ug/dl, 24. 9 ug/dl accompanied by a data flag, and 21. 6 ug/dl. The sample was tested at another laboratory on a cobas 6000 e 601 module, resulting with a t4 value of 12. 21 ug/dl. No adverse events were alleged to have occurred with the patient. The serial number of the customer's e 801 analyzer is (b)(4). The serial number of the e 601 analyzer used at the other laboratory was asked for, but not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-03894
MDR Report Key8018332
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-10-30
Date of Report2018-12-27
Date of Event2018-10-09
Date Mfgr Received2018-10-09
Date Added to Maude2018-10-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS T4 ASSAY
Generic NameRADIOIMMUNOASSAY, TOTAL THYROXINE
Product CodeCDX
Date Received2018-10-30
Model NumberNA
Catalog Number07027885190
Lot Number287093
Device Expiration Date2019-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.