VETPRO VETPRO DC 2.0 DCV7-G2A/H

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-10-30 for VETPRO VETPRO DC 2.0 DCV7-G2A/H manufactured by Midmark Corporation.

Event Text Entries

[126553095] On (b)(6) 2018 it was reported to midmark by a dealer technician that an x-ray unit (vetpro dc 2. 0, 76" reach, w/two stud mount; serial number (b)(4)) separated from the wall in which it was secured. There were no injuries or contact with any person. The unit detached from the wall when the facility was closed and discovered in the morning by facility staff, as reported by the complainant. The unit mounting was visually inspected at the user facility by the technician and photographs were received by midmark technical services. The technician, after inspection of the mounting apparatus concluded that the bolts were not properly installed by the installation technician at the time of unit install. The lag bolts, provided by midmark, were installed into the wall and did not correctly connect with the stud causing the machine to not be securely attached to the wall. The original hardware provided with the x-ray system was observed to be present but the lag screws were mounted off-center on the stud or completely missed the stud. It is unknown if the installation manual was used at the time of initial installation. No further information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1423380-2018-00004
MDR Report Key8018741
Date Received2018-10-30
Date of Report2018-10-29
Date of Event2018-10-17
Date Mfgr Received2018-10-17
Date Added to Maude2018-10-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactADAM FORESMAN
Manufacturer Street675 HEATHROW DRIVE
Manufacturer CityLINCOLNSHIRE IL 60069
Manufacturer CountryUS
Manufacturer Postal60069
Manufacturer Phone2242204329
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVETPRO
Generic NameVETPRO DC
Product CodeEAP
Date Received2018-10-30
Model NumberVETPRO DC 2.0
Catalog NumberDCV7-G2A/H
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMIDMARK CORPORATION
Manufacturer Address675 HEATHROW DRIVE LINCOLNSHIRE IL 60069 US 60069


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-30

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.