ERGOJUST CART IP-HD-EC-VIDEO-EXT N/A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-10-30 for ERGOJUST CART IP-HD-EC-VIDEO-EXT N/A manufactured by Natus Medical Inc. Dba Excle-tech Ltd (xltek).

Event Text Entries

[125618276] This is a known issue which has been previously investigated and resulted in a fca. The customer has been informed about the field corrective action. The retrofit kit has been provided to customer in august to install on the cart to prevent any such failure. The customer did not installed the kit on the cart. The upgraded version of replacement ergo just cart has been shipped to the customer.
Patient Sequence No: 1, Text Type: N, H10


[125618277] Ergojust video extension broke where it mounts to the base of the cart. Pictures that the customer sent, shows it broke at the welds. There was no patient or user injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612330-2018-00016
MDR Report Key8019648
Report SourceHEALTH PROFESSIONAL
Date Received2018-10-30
Date of Report2018-10-19
Date of Event2018-10-08
Date Mfgr Received2018-10-19
Date Added to Maude2018-10-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR SANJAY MEHTA
Manufacturer Street2568 BRISTOL CIRCLE
Manufacturer CityOAKVILLE, ONTARIO L6H5S1
Manufacturer CountryCA
Manufacturer PostalL6H5S1
Manufacturer G1NATUS MEDICAL INCORPORATED DBA EXCEL-TECH (XLTEK)
Manufacturer Street2568 BRISTOL CIRCLE
Manufacturer CityOAKVILLE, ONTARIO L6H5S1
Manufacturer CountryCA
Manufacturer Postal CodeL6H5S1
Single Use3
Remedial ActionRC
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameERGOJUST CART
Generic NameELECTROENCEPHALOGRAPH
Product CodeGWQ
Date Received2018-10-30
Model NumberIP-HD-EC-VIDEO-EXT
Catalog NumberN/A
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INC. DBA EXCLE-TECH LTD (XLTEK)
Manufacturer Address2568 BRISTOL CIRCLE OAKVILLE, OAKVILLE ONTARIO L6H 5S1 CA L6H 5S1


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-10-30

© 2026 FDA.report