FLEXOR PARALLEL URETERAL ACCESS SHEATH AND DILATORS FUS-107045-P

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-10-30 for FLEXOR PARALLEL URETERAL ACCESS SHEATH AND DILATORS FUS-107045-P manufactured by Cook Inc.

Event Text Entries

[125611129] (b)(6). (b)(4). This report includes information known at this time. A follow up report will be submitted should additional relevant information become available.
Patient Sequence No: 1, Text Type: N, H10


[125611130] It was initially reported, in a flexor parallel a little piece of coating detached during the procedure. Additional information was received on (b)(6) 2018. During a retrograde intrarenal surgery (rirs) right kidney, the stone was very hard and during the procedure with extractor and stone inside, the stone scratched the flexor parallel ureteral access sheath coating and a little slice went into the kidney. The doctor was able to remove this little piece of coating with the extractor. No additional procedure was required and there were no adverse effects to the patient due to this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2018-03131
MDR Report Key8020209
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-10-30
Date of Report2018-11-27
Date of Event2018-10-09
Date Mfgr Received2018-11-16
Date Added to Maude2018-10-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. LARRY POOL
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFLEXOR PARALLEL URETERAL ACCESS SHEATH AND DILATORS
Generic NameKOE DILATOR, URETHRAL
Product CodeKOE
Date Received2018-10-30
Returned To Mfg2018-11-12
Catalog NumberFUS-107045-P
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-30

© 2026 FDA.report