DRX-REVOLUTION MOBILE X-RAY SYSTEM DRX-R-1

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-10-31 for DRX-REVOLUTION MOBILE X-RAY SYSTEM DRX-R-1 manufactured by Carestream Health, Inc.

Event Text Entries

[125639015] An x-ray tech needed to position a portable x-ray unit above the patient. Prior to this, a radiant warmer that was providing therapy to the patient had to be repositioned away from above the patient to make room for the portable x-ray. The radiant warmer was moved away from above the patient, but not enough to clear the counterbalance of the x-ray unit. When the x-ray tech moved the collimator down toward the patient, the x-ray counterbalance moved up, raising the radiant warmer attached to the patient's bed. Eventually, the radiant warmer slipped off of the counterbalance and the bed drops to the ground. Approximately 4 or 5 inches. The x-ray technician did not move the radiant warmer arm, a nurse did. The x-ray unit has no indication that the counterbalance raises up in the air. That knowledge is only known by experience. This probably contributed to this event. No patient harm, but considering the acuity of the patient, this was truly a "near miss".
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8023174
MDR Report Key8023174
Date Received2018-10-31
Date of Report2018-10-29
Date of Event2018-10-28
Report Date2018-10-29
Date Reported to FDA2018-10-29
Date Reported to Mfgr2018-10-31
Date Added to Maude2018-10-31
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDRX-REVOLUTION MOBILE X-RAY SYSTEM
Generic NameSYSTEM, X-RAY, MOBILE
Product CodeIZL
Date Received2018-10-31
Model NumberDRX-R-1
Device AvailabilityY
Device Age2 YR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCARESTREAM HEALTH, INC
Manufacturer Address150 VERONA ST ROCHESTER NY 14608 US 14608


Patients

Patient NumberTreatmentOutcomeDate
10 2018-10-31

© 2026 FDA.report