ACL TOP 500 CTS 2800-40 00000280040

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2018-10-31 for ACL TOP 500 CTS 2800-40 00000280040 manufactured by Instrumentation Laboratory.

Event Text Entries

[125650474] An investigation was conducted that included a review of the instrument clot curves. All aptt results illustrated well-formed curve formation, with appropriate clotting times detected and no concerning errors or warnings. The conclusion is that the values obtained on the acl top 500 cts were appropriately determined by the analyzer and the instrument was performing as intended. Based on the above, there was no indication of an instrument malfunction and no remedial action is needed.
Patient Sequence No: 1, Text Type: N, H10


[125650475] A customer reported that their acl top 500 cts generated an erroneous aptt result (36. 9 seconds) using hemosil synthasil. A patient was administered an increased dose of heparin based on this erroneous aptt result. Subsequently, another sample was drawn from the same patient (approximately 3 hours later) and it reported an aptt result of > 200. The heparin administration was stopped and the customer indicated that the patient's aptt result was brought back into the therapeutic range. There were no further reports of clinical implications for the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1217183-2018-00009
MDR Report Key8024101
Report SourceCOMPANY REPRESENTATIVE
Date Received2018-10-31
Date of Report2018-10-31
Date of Event2018-10-11
Date Mfgr Received2018-10-12
Date Added to Maude2018-10-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. REBA DAOUST
Manufacturer Street180 HARTWELL ROAD
Manufacturer CityBEDFORD MA 01730
Manufacturer CountryUS
Manufacturer Postal01730
Manufacturer Phone7818614597
Manufacturer G1INSTRUMENTATION LABORATORY
Manufacturer Street180 HARTWELL ROAD
Manufacturer CityBEDFORD MA 01730
Manufacturer CountryUS
Manufacturer Postal Code01730
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACL TOP 500 CTS
Generic NameACL TOP, PRODUCT CODE: GKP
Product CodeGKP
Date Received2018-10-31
Model Number2800-40
Catalog Number00000280040
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINSTRUMENTATION LABORATORY
Manufacturer Address180 HARTWELL ROAD BEDFORD MA 01730 US 01730


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-10-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.