FACE PLATE ASSEMBLY O-X-L VA-LAB- 100 UL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-10-30 for FACE PLATE ASSEMBLY O-X-L VA-LAB- 100 UL manufactured by Amico Patient Care Corporation.

Event Text Entries

[126007428] Pt came to the ed with suspected aspiration pneumonia. Pt was intubated and going to be transferred to the icu eventually. The staff was going to connect the pt to the wall oxygen and the oxygen flow meter bent forward. The pt desaturated and the staff had to retrieve an oxygen tank. The pt's heart rate slowed. The pt received cpr, epinephrine, and rosc was returned. After several days, the pt recovered and was discharged. The facilities tech replaced the face plate. The tech thinks that the inner piece of the plate was broken. The broken plate was not available to be assessed. The plate was inadvertently discarded by someone. The info in this report was obtained from a new face plate. The report is based off the premise that the broken face plate was made by the same mfr as the new face plate.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5080981
MDR Report Key8028252
Date Received2018-10-30
Date of Report2018-10-26
Date of Event2018-09-07
Date Added to Maude2018-11-01
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameFACE PLATE ASSEMBLY
Generic NameFLOWMETER, TUBE, THORPE, BACK-PRESSURE
Product CodeCAX
Date Received2018-10-30
Model NumberO-X-L VA-LAB- 100 UL
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerAMICO PATIENT CARE CORPORATION
Manufacturer Address85 FULTON WAY RICHMOND HILL, ON L4B 2N4 CA L4B 2N4


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-10-30