ETEST? IMIPENEM IP 32WW S30 412374

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-11-01 for ETEST? IMIPENEM IP 32WW S30 412374 manufactured by Biomerieux, Sa.

Event Text Entries

[127360116] A customer in (b)(6) reported false resistant imipenem results for pseudomonas aeruginosa isolates in association with etest? Imipenem ip 32ww s30 (lot 1006022690). In (b)(6) 2018, the customer initially reported false results with vitek? Cards, when performing a study for a diploma thesis, on different techniques with ast testing. The customer stated the initial test was with a patient urine sample on a vitek 2 ast-n248 card, and that other tests for the study included the vitek 2 ast-n196, bd phoenix, etest and disk diffusion. The investigation for the vitek cards, concluded that the customer's etest results were false resistant compared to broth microdilution. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00227
MDR Report Key8030587
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-11-01
Date of Report2018-11-01
Date Mfgr Received2018-10-16
Device Manufacturer Date2017-10-12
Date Added to Maude2018-11-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DEBRA BROYLES
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOMERIEUX, SA
Manufacturer Street3 ROUTE DE PORT MICHAUD
Manufacturer CityLA BALME, LES GROTTES ISERE 38390
Manufacturer CountryFR
Manufacturer Postal Code38390
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameETEST? IMIPENEM IP 32WW S30
Generic NameETEST? IMIPENEM IP 32WW S30
Product CodeJWY
Date Received2018-11-01
Catalog Number412374
Lot Number1006022690
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, SA
Manufacturer Address3 ROUTE DE PORT MICHAUD LA BALME, LES GROTTES ISERE 38390 FR 38390