3S HEMI IMPLANT, SMALL, NON-STERILE 101-00-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-11-02 for 3S HEMI IMPLANT, SMALL, NON-STERILE 101-00-001 manufactured by Trilliant Surgical.

Event Text Entries

[126067938] The event is regarding a 3s hemi small implant originally implanted in 2013 that malfunctioned. The implant was noted by the patient in (b)(6) 2017 to have become visible on the dorsal aspect of the surgical site. As a result, the surgeon reviewed the implant in (b)(6) 2018, noted the implant was in excellent placement, yet removed it in (b)(6) 2018. The surgeon noted there is a hole present on the dorsal aspect of the bone, post removal. There is no information regarding an additional surgery to correct any bone interference at this time. The complaint was confirmed by review of the provided x-ray. As discussed in the surgical technique review section, there are several contributing factors that could have resulted in the reported event; however, none can be definitively determined due to limited details provided. Review of the complaint log revealed one other similar event. Based on the above investigation, no other actions is deemed necessary at this time. If additional details that would impact the conclusion of this report are reported to trilliant surgical, a supplemental report shall be submitted within the defined timeframe.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007420745-2018-00017
MDR Report Key8035493
Date Received2018-11-02
Date of Report2018-11-02
Date of Event2018-09-01
Date Mfgr Received2018-10-05
Date Added to Maude2018-11-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. JODIE RODRIGUEZ
Manufacturer Street727 N SHEPHERD DR SUITE 100
Manufacturer CityHOUSTON TX 77007
Manufacturer CountryUS
Manufacturer Postal77007
Manufacturer Phone8004952919
Manufacturer G1TRILLIANT SURGICAL
Manufacturer Street727 N SHEPHERD DR SUITE 100
Manufacturer CityHOUSTON TX 77007
Manufacturer CountryUS
Manufacturer Postal Code77007
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name3S HEMI IMPLANT, SMALL, NON-STERILE
Generic Name3S HEMI IMPLANT, SMALL, NON-STERILE
Product CodeKWD
Date Received2018-11-02
Returned To Mfg2018-10-10
Model Number101-00-001
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTRILLIANT SURGICAL
Manufacturer Address727 N SHEPHERD DR SUITE 100 HOUSTON 77007 US 77007


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2018-11-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.