ARCHITECT LH 6C25-27

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2007-01-04 for ARCHITECT LH 6C25-27 manufactured by Abbott Laboratories.

Event Text Entries

[16398604] The customer states that non-abbott controls are reading high out of specifications with the architect i2000 lh assay (lot 33198m200). There is no impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5


[16482982] This issue was identified while investigating an increase in customer complaints for upward shifts in non-abbott control and patient sample results, as well as calibration errors. Specifically, error code 1227 "assay (lh) number (641) calibration failure, fit concentration too high for cal f" may be generated which would result in failure to obtain a valid calibration curve. Abbott has not received any reports of adverse events related to the affected lots of architect lh reagents. An investigation is in process.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1415939-2007-00001
MDR Report Key803731
Report Source01,05,06
Date Received2007-01-04
Date of Report2007-01-03
Date of Event2006-09-04
Date Mfgr Received2007-01-03
Device Manufacturer Date2005-12-01
Date Added to Maude2007-01-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactDR. DAVID BARCH, D-09Y6, AP6C-2
Manufacturer Street100 ABBOTT PARK ROAD
Manufacturer CityABBOTT PARK IL 600646092
Manufacturer CountryUS
Manufacturer Postal600646092
Manufacturer Phone8479379328
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number1415939-11/29/06-004-C
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT LH
Generic NameCMIA FOR THE QUANTITATIVE DETERMINATION OF HUMAN LUTEINIZING
Product CodeCEP
Date Received2007-01-04
Model NumberNA
Catalog Number6C25-27
Lot Number33198M200
ID NumberNA
Device Expiration Date2007-06-16
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key791232
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK ROAD ABBOTT PARK IL 600643500 US
Baseline Brand NameARCHITECT LH
Baseline Generic NameCMIA FOR THE QUANTITATIVE DETERMINATION OF HUMAN LUTEINIZING
Baseline Model NoNA
Baseline Catalog No6C25-27
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 2007-01-04

© 2026 FDA.report