30 -70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CF-250

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-11-05 for 30 -70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CF-250 manufactured by Charter Medical, Ltd..

Event Text Entries

[127866517] The cf-250 device has two spike port components. The spike port component contains a septum membrane in the interior tube. The cf-250 is 100% leak tested during production. The returned bag was visually inspected; both spike port components of the bag contained spike components. The returned bag had an area of interest identified on one of the spike port components of the bag. Two observations were noted: a tear/hole was verified on one of the spike ports. The hole occurred in the side of the spike port tube, below the septum. A hole at this location would have been detected upon filling if it existed at that time (i. E. Before a spike would have been inserted into the spike port). The hole was reported after the bag had been frozen and thawed. Additionally, a hole in this location would likely be detected during production leak test if it existed at that time. The remnants of the spike port septum membrane were observed inside the bag. This observation is likely an indication of excessive force or incorrect user technique during the spiking process. The location and nature of the hole in the spike port component, along with the process step (after thawing and hanging the bag) when the leak was detected, suggests user excessive force or spiking technique likely contributed to the hole on the returned sample.
Patient Sequence No: 1, Text Type: N, H10


[127866518] A leak, from the port component of a single cf-250 product was reported by the user facility. The bag was thawed for infusion and hung. A hole was observed in the port component.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2018-00018
MDR Report Key8040144
Date Received2018-11-05
Date of Report2018-11-05
Date of Event2018-09-12
Date Mfgr Received2018-10-04
Device Manufacturer Date2016-08-12
Date Added to Maude2018-11-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. TODD MEINECKE
Manufacturer Street3948-A WESTPOINT BLVD
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name30 -70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Generic Name30 -70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Product CodeLPZ
Date Received2018-11-05
Returned To Mfg2018-10-12
Model NumberCF-250
Catalog NumberCF-250
Lot Number145979
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-05

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.