[126635958] Quality control (qc) results were unacceptable after installation of multiple lots of roche diagnostics cobas chloride electrodes (cat # 03246353001). Roche sent a product bulletin on (b)(6) 2018 (tp-00394; chloride electrodes - discontinue use of certain lots). However, we have observed this qc issue on numerous lots not listed in the product bulletin. Chloride electrode lots which did not pass qc criteria at our institute include: a2072, b0625, b0094, a3760, w5233, x1881, x1882, x6230, y7004, y7200, and y8413. Affected electrodes were not placed into clinical use by our lab, therefore no pt results have been affected at our facility.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5081091
MDR Report Key8042449
Date Received2018-11-05
Date of Report2018-11-02
Date Added to Maude2018-11-06
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Sequence Number: 1

Brand NameCHLORIDE ELECTRODE
Generic NameELECTRODE, ION-SPECIFIC, CHLORIDE
Product CodeCGZ
Date Received2018-11-05
Catalog Number03246353001
Lot NumberA2072
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS OPERATIONS INC.
Manufacturer AddressINDIANAPOLIS IN 46250 US 46250

Device Sequence Number: 2

Brand NameCHLORIDE ELECTRODE
Generic NameELECTRODE, ION-SPECIFIC, CHLORIDE
Product CodeCGZ
Date Received2018-11-05
Catalog Number03246353001
Lot NumberB0625
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No2
Device Event Key0
ManufacturerROCHE DIAGNOSTICS OPERATIONS INC.
Manufacturer AddressINDIANAPOLIS IN 46250 US 46250

Device Sequence Number: 3

Brand NameCHLORIDE ELECTRODE
Generic NameELECTRODE, ION-SPECIFIC, CHLORIDE
Product CodeCGZ
Date Received2018-11-05
Catalog Number03246353001
Lot NumberB0094
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No3
Device Event Key0
ManufacturerROCHE DIAGNOSTICS OPERATIONS INC.
Manufacturer AddressINDIANAPOLIS IN 46250 US 46250

Device Sequence Number: 4

Brand NameCHLORIDE ELECTRODE
Generic NameELECTRODE, ION-SPECIFIC, CHLORIDE
Product CodeCGZ
Date Received2018-11-05
Catalog Number03246353001
Lot NumberA3760
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No4
Device Event Key0
ManufacturerROCHE DIAGNOSTICS OPERATIONS, INC.
Manufacturer AddressINDIANAPOLIS IN 46250 US 46250

Device Sequence Number: 5

Brand NameCHLORIDE ELECTRODE
Generic NameELECTRODE, ION-SPECIFIC, CHLORIDE
Product CodeCGZ
Date Received2018-11-05
Catalog Number03246353001
Lot NumberW5233
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No5
Device Event Key0
ManufacturerROCHE DIAGNOSTICS OPERATIONS, INC.
Manufacturer AddressINDIANAPOLIS IN 46250 US 46250

Device Sequence Number: 6

Brand NameCHLORIDE ELECTRODE
Generic NameELECTRODE, ION-SPECIFIC, CHLORIDE
Product CodeCGZ
Date Received2018-11-05
Catalog Number03246353001
Lot NumberX1881
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No6
Device Event Key0
ManufacturerROCHE DIAGNOSTICS OPERATIONS, INC.
Manufacturer AddressINDIANAPOLIS IN 46250 US 46250

Device Sequence Number: 7

Brand NameCHLORIDE ELECTRODE
Generic NameELECTRODE, ION-SPECIFIC, CHLORIDE
Product CodeCGZ
Date Received2018-11-05
Catalog Number03246353001
Lot NumberX1882
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No7
Device Event Key0
ManufacturerROCHE DIAGNOSTICS OPERATIONS, INC.
Manufacturer AddressINDIANAPOLIS IN 46250 US 46250

Device Sequence Number: 8

Brand NameCHLORIDE ELECTRODE
Generic NameELECTRODE, ION-SPECIFIC, CHLORIDE
Product CodeCGZ
Date Received2018-11-05
Catalog Number03246353001
Lot NumberX6230
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No8
Device Event Key0
ManufacturerROCHE DIAGNOSTICS OPERATIONS, INC.
Manufacturer AddressINDIANAPOLIS IN 46250 US 46250

Device Sequence Number: 9

Brand NameCHLORIDE ELECTRODE
Generic NameELECTRODE, ION-SPECIFIC, CHLORIDE
Product CodeCGZ
Date Received2018-11-05
Catalog Number03246353001
Lot NumberY7004
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No9
Device Event Key0
ManufacturerROCHE DIAGNOSTICS OPERATIONS, INC.
Manufacturer AddressINDIANAPOLIS IN 46250 US 46250

Device Sequence Number: 10

Brand NameCHLORIDE ELECTRODE
Generic NameELECTRODE, ION-SPECIFIC, CHLORIDE
Product CodeCGZ
Date Received2018-11-05
Catalog Number03246353001
Lot NumberY7200
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No10
Device Event Key0
ManufacturerROCHE DIAGNOSTICS OPERATIONS, INC.
Manufacturer AddressINDIANAPOLIS IN 46250 US 46250

Device Sequence Number: 11

Brand NameCHLORIDE ELECTRODE
Generic NameELECTRODE, ION-SPECIFIC, CHLORIDE
Product CodeCGZ
Date Received2018-11-05
Catalog Number03246353001
Lot NumberY8413
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No11
Device Event Key0
ManufacturerROCHE DIAGNOSTICS OPERATIONS, INC.
Manufacturer AddressINDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.