GLUC2 GLUCOSE HK 20767131322

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-11-07 for GLUC2 GLUCOSE HK 20767131322 manufactured by Roche Diagnostics.

Event Text Entries

[128259801] This event occurred in (b)(6). Phone was provided as (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[128259802] The customer received questionable low gluc3 glucose hk gen. 2 results for one patient from two cobas 6000 c501 modules. Only one serial number ((b)(4)) was provided. The patient was not responsive at the dentist and when the ambulance arrived, the patient became responsive again but was taken to hospital. At the hospital, for the first sample the glucose results from two different cobas 6000 analyzers were 11 mg/dl and 4 mg/dl. The result from a bga system for a different tube from for the same draw was 90 mg/dl. For a second sample, the glucose results from two different cobas 6000 analyzers were 47 mg/dl and 43 mg/dl. The result from a bga system for a different tube from for the same draw was 102 mg/dl. The result from a poct- bz system was 89 mg/dl. For a third sample, the glucose results from two different cobas 6000 analyzers 93 mg/dl and 95 mg/dl. All of the results were reported outside of the laboratory. The patient was not adversely affected.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-04120
MDR Report Key8045772
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-11-07
Date of Report2019-01-04
Date of Event2018-10-17
Date Mfgr Received2018-10-19
Date Added to Maude2018-11-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameGLUC2 GLUCOSE HK
Generic NameHEXOKINASE, GLUCOSE
Product CodeCFR
Date Received2018-11-07
Model NumberNA
Catalog Number20767131322
Lot Number336546
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-07

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.