4.2MM RADIOLUCENT DRILL BIT 4200355042

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-11-07 for 4.2MM RADIOLUCENT DRILL BIT 4200355042 manufactured by Stryker Instruments-kalamazoo.

Event Text Entries

[126430768] The device is available for return. A follow up report will be filed once the quality investigation is complete.
Patient Sequence No: 1, Text Type: N, H10


[126430769] It was reported that during a osteosynthesis for femoral condylar fracture surgery; the drill bit tip broke. It was further reported that there were no delays and no adverse consequences as a result of this event and the procedure was completed successfully.
Patient Sequence No: 1, Text Type: D, B5


[131836283] Investigation results indicate that contact metal while in use, caused damage to the cutting surface and resulted in a break at the proximal end of the shaft. The associated ifu contains the following caution "caution: excessive pressure can damage either the drill bit or driver. " the quality investigation is complete.
Patient Sequence No: 1, Text Type: N, H10


[131836284] It was reported that during a osteosynthesis for femoral condylar fracture surgery; the drill bit tip broke. It was further reported that there were no delays and no adverse consequences as a result of this event and the procedure was completed successfully.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001811755-2018-02384
MDR Report Key8047654
Report SourceHEALTH PROFESSIONAL
Date Received2018-11-07
Date of Report2018-12-21
Date of Event2018-10-10
Date Mfgr Received2018-12-21
Date Added to Maude2018-11-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. UNA BARRY
Manufacturer StreetINSTRUMENTS DIVISION CARRIGTWOHILL BUS. & TECH PARK
Manufacturer CityCARRIGTWOHILL NA
Manufacturer PostalNA
Manufacturer Phone214532900
Manufacturer G1STRYKER INSTRUMENTS-IRELAND
Manufacturer StreetINSTRUMENTS DIVISION CARRIGTWOHILL BUS. & TECH PARK
Manufacturer CityCARRIGTWOHILL NA
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name4.2MM RADIOLUCENT DRILL BIT
Generic NameBIT, SURGICAL
Product CodeGFG
Date Received2018-11-07
Returned To Mfg2018-11-13
Catalog Number4200355042
Lot Number18135017
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER INSTRUMENTS-KALAMAZOO
Manufacturer Address4100 EAST MILHAM AVENUE KALAMAZOO MI 49001 US 49001


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.