ALGO 5 NEWBORN HEARING SCREENER 000619

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-11-07 for ALGO 5 NEWBORN HEARING SCREENER 000619 manufactured by Natus Medical Incorporated.

Event Text Entries

[128589053] The entire algo5 system was requested for return to (b)(6) for evaluation and a replacement algo 5 system was provided to customer to prevent delay in screening. The device was received by (b)(6) medical for evaluation on (b)(6) 2017 and is under evaluation. If additional information is available (b)(6) medical will file a supplemental.
Patient Sequence No: 1, Text Type: N, H10


[128589054] (b)(6) medical received a complaint on (b)(6) 2018. It was reported to (b)(6) that algo 5 hearing screener results were questionable to the user. Customer stated that patient passed hearing screen (left: pass, right: refer) but was later diagnosed as deaf and confirmed that equipment checks had passed prior to screening. There was no death, serious injury, delay in treatment, or environmental/safety concerns. (b)(6) technical service representative suggested customer to discontinue use of the device and requested answers to additional questions regarding the event. After the initial follow up with customer by (b)(6) technical service representative, it was confirmed that the screening was performed while the patient was sleeping and there was no sensor replacement or adhesion issues. The follow up confirmatory tested was completed on the same day as screen and the patient was diagnosed with bilateral hearing loss. The patients genetic testing indicated that parents have gene for hearing loss and the patient's older sister was born deaf.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2018-00151
MDR Report Key8049144
Report SourceUSER FACILITY
Date Received2018-11-07
Date of Report2018-10-08
Date of Event2018-10-02
Date Mfgr Received2018-11-12
Device Manufacturer Date2016-08-31
Date Added to Maude2018-11-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES FITZGERALD
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685173
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameALGO 5 NEWBORN HEARING SCREENER
Generic NameALGO 5
Product CodeGWJ
Date Received2018-11-07
Returned To Mfg2018-11-05
Model Number000619
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
10 2018-11-07

© 2026 FDA.report