AIRLIFE? INFANT SKIN TEMPERATURE PROBE, SINGLE PATIENT USE (GIRAFFE?, PANDA?) 2074816-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-11-08 for AIRLIFE? INFANT SKIN TEMPERATURE PROBE, SINGLE PATIENT USE (GIRAFFE?, PANDA?) 2074816-001 manufactured by Vyaire Medical.

Event Text Entries

[126517178] Vyaire complaint (b)(4): any additional information provided by the customer will be included in a follow-up report. At this time, vyaire has not received the suspect device/component for evaluation.
Patient Sequence No: 1, Text Type: N, H10


[126517179] It was reported to vyaire that a nurse was repositioning the temperature-probe cover, she lifted off the cover (hydrogel heart) and discovered that the temperature probe sensor had embedded into the skin of a neonate. The infant was reported to have sustained a skin tear. Area of erythema 0. 5cm x0. 5 cm noted over right mid abdomen due to temp probe burn; placed hydrogel dressing over the area and lesion is improving, no jaundice.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012348048-2018-00001
MDR Report Key8050651
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-11-08
Date of Report2018-10-11
Date of Event2018-08-31
Date Mfgr Received2018-10-11
Date Added to Maude2018-11-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STANLEY TAN
Manufacturer Street22745 SAVI RANCH PKWY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer Phone7149193324
Manufacturer G1AMPHENOL (CHANGZHOU) CONNECTOR SYSTEMS CO., LTD.
Manufacturer StreetNO. 8 XIHU ROADWUJIN HI THE DEVELOPMENT ZONE CHANGZHOU
Manufacturer CityJIANGSU, 213164
Manufacturer CountryCH
Manufacturer Postal Code213164
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAIRLIFE? INFANT SKIN TEMPERATURE PROBE, SINGLE PATIENT USE (GIRAFFE?, PANDA?)
Generic NameINCUBATOR, NEONATAL
Product CodeFMZ
Date Received2018-11-08
Catalog Number2074816-001
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL
Manufacturer Address22745 SAVI RANCH PKWY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-11-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.